2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 465
TMP- 3.6.0 PDIA MORE THAN 6.0 / ISOLATE TRANSDUCER #9 / POSSIBLE OPEN
a) A
known good
1
pressure sensor, from your spare parts, is required to isolate the dialysate pressure
circuit from the sensor circuit.
3
To ENSURE the sensor is
known good
, when it is plugged into distribution board position X9,
“PDIAL”, on ANOTHER machine that is not exhibiting TMP problems, it will drive the machine’s
debug screen 10
PDIA
to between 3.5 and 6.0.
b) Figure right, plug the
known good
pressure sensor
into the malfunctioning machine’s distribution board
position X9, “PDIAL”. There is NO NEED to install the
pressure sensor into the hydraulic yet.
c) From the malfunctioning machine’s debug screen 10, is
PDIA
(middle column) between 3.5 and 6.0?
Yes
PDIA
between 3.5 and 6.0! The malfunctioning machine’s Dialysate Pressure Transducer
#9 is bad.
No
PDIA
IS NOT between 3.5 and 6.0! See parts A through D below:
A)
Turn the machine OFF.
B)
Swap in the following components (see
COMPONENT LIST
below), one at a time, with
known good, and in between continue to parts C and D to see if the new component fixed
the problem.
COMPONENT LIST: 1)
Actuator-Test Board
2
;
2)
Sensor Board
1
;
3)
Sensor Board cable;
4)
Functional Board
1, 2
;
5)
Distribution board;
6)
Motherboard
1
1
To locate the boards refer to
(page 9)
2
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
)
C)
With the known good pressure sensor remaining plugged in, return to Dialysis Program.
D)
If screen 10’s
PDIA
is still NOT between 3.5 and 6.0 return to part A.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...