2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 318
L) Call the Home screen. WITHOUT LOOKING AWAY, watch [
Temperature
]
AND [
Conductivity
] for one (1) minute. Based first on
[
Temperature
]
,
TWO (2) possible scenarios below
1)
IF between 35.5 and 38.5° C AND STABLE i.e. does NOT changes
more than +/- 0.2 °C:
Going forward, if [
Temperature
] becomes
unstable address it FIRST! For now see part M.
2) ALL OTHERS:
Proceed to
page 214
SECTION 4 – TEMPERATURE PROBLEMS
M) Based on
[Conductivity]
, TWO (2) possible scenarios:
1)
IF (and ONLY if) between 13.0 and 14.3 mS AND STABLE i.e. does NOT change more than
+/- 0.2:
See
(page 319).
2)
ALL OTHERS:
See part N.
N)
Firmly connect to JUGS* of known good** bicarb AND acid!
*
To isolate between a Central Concentrate Delivery System (SDS) and the 2008 K/K
2
connect to
JUGS!
**
Tested good per clinic procedures!
O) ENSURE [Dialysate Flow] is ON (Dialysate Flow on/off lamp is NOT blinking)!
R) If the acid AND / OR bicarb source was changed above allow five (5) minutes BEFORE continuing to
part S.
(page 319).