2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 335
CO- 1.0.9 VERIFY DIALYSATE LIMITS
a) Select the ‘Dialysate’ tab (Figure below).
Dialysate SCREEN
b)
TCD
is based on Dialysate Composition and indicates what the ‘Actual’ Conductivity would be if
everything was perfect.
c) Using [Alarm Width], spread Conductivity Limits as wide as they will go and press ‘CONFIRM’.
d) Using [Alarm Position], set the upper Limit to 0.5 above
TCD
and press ‘CONFIRM’.
e) Based on the ‘Actual’ [
Conductivity
] window (Figure right),
TWO (2) possible scenarios:
1)
IF (and ONLY if) [Conductivity] is STABLE between 13.4
and 14.3 mS:
(page 336).
2) ALL OTHERS:
Proceed to
page 337
, procedure number