2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 337
CO- 2.0.0 VERIFY PRIMARY TEMPERATURE ALARM LIMITS
a) Call the Home screen.
b) ENSURING [
Temperature
] REMAINS between 35.5 and 38.5° C,
what color is its window (Figure right)?
1)
IF pale yellow/white:
Proceed to
page 337
, procedure number
2)
IF RED:
Proceed to
page 222
, procedure number
CO- 2.0.6 VERIFY PRIMARY CONDUCTIVITY ALARM LIMITS
Call the Home screen. What color is the [
Conductivity
] window (Figure
right)? TWO (2) possible scenarios:
1)
IF (and ONLY if) pale yellow / white:
page 340
2) IF RED:
See parts a AND b below:
a) At the bottom of the screen, select the ‘Dialysate’ tab.
b) Figure right, using [Alarm Position], can the Limits be
adjusted so that ‘Actual’
Conductivity
is
CENTERED between them?
Yes
‘Actual’ Conductivity between the Limits!
Proceed to
page 339
, procedure number
No Limits cannot be adjusted properly! See procedure number
(page 337).
CO- 2.0.61 [CONDUCTIVITY] WINDOW IS RED
a) Figure right, place the
274 Ω
resistor plug, from the
TWO-
RESISTOR SET
, into distribution
board position, “X7, COND”.
b) Can the Limits now be adjusted so
that ‘Actual’
Conductivity
is
between them?
Yes
‘Actual’ Conductivity between the Limits! See procedure number
(page 339).
No The Limits CANNOT be adjusted properly! See parts a THROUGH d below:
a) Return Conductivity Cell #7’s connector to distribution board position “X7, COND”.
b) Return the concentrate connectors and close the shunt door.
Parts c and d next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...