2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 145
NW- 1.0.8 IN HEAT DISINFFECT “NO WATER” REOCCURS
a) Call debug screen 4. If debug does not appear press ‘Escape’ then call screen 4.
b) When Valve #43 is closed
*
,
PDIA
(left column)
NORMALLY
cycles between 0.0
and to more than 0.1. When Valve #43 opens it stays more than 1.0 but
NORMALLY for no more than sixteen (16) seconds.
*
In Heat Disinfect Valve #43 automatically cycles between closed and open
c) WITHOUT LOOKING AWAY, watch
PDIA
for forty five (45) seconds. If it remains more than 0.1
continue to watch for forty five (45) more seconds. Does it EVER cycle to 0.0?
Yes
PDIA
cycles to 0.0! See procedure number
page 146
No
PDIA
REMAINS ALWAYS more than 0.1! See parts a AND b below:
a)
CAREFUL HERE! Per the Figure below
double clamp Valve #43’s
OUTPUT tubing.
NOTE:
Valve #43’s output tubing extends towards the front of the
machine!
b) Watch
PDIA
again for forty five (45) seconds. Does it NOW cycle to 0.0?
Yes
PDIA
now cycles to 0.0!
Remove the clamps!
TWO (2) possible bad
components:
1)
Valve #43;
2)
Actuator-Test Board.
No
PDIA
still DOES NOT cycle to 0.0!
A)
ENSURE Valve #43 is double
clamped properly!
B)
Return to Dialysis Program, allow eight (8) minutes,
and then check for TMP problems.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...