2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 660
DVF- 5.0.0 PIN 43 (VALVE #38) LESS THAN 40
Ω
a)
Figure below
, unplug
Valve #38
from the distribution board.
b) Using a flashlight, check inside the vacant distribution board position.
If corrosion or damage is
located the distribution board needs to be replaced!
c)
Leave Valve #38 unplugged
until instructed
.
d) Turn the machine on AND
return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
e) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
f) Call debug screen 0 then allow thirty (30) seconds.
g) Do the balancing chamber ‘dots’ begin to cycle between blue and white?
Yes
The balancing chamber ‘dots’ are cycling! Balancing Chamber Valve #38
*
OR its blue wire
harness is bad!
*
To LOCATE Valve #38 refer to
(page 193).
No
a) Return Valve #38 to its distribution board position
b) See (ABOVE) procedure number
(page 654).