2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 225
T- 1.0.130 [TEMPERATURE] WINDOW IS PALE/YELLOW
a) Figure right, at the distribution board, unplug the
2
nd
connector from the LEFT, “MON-NTC”! This is temp
sensor NTC #3.
b) Based on the [
Temperature
] window, TWO (2) possible
scenarios:
1)
IF (and ONLY if) LESS THAN 30.0° C:
(page 225).
2) IF MORE THAN 30° C:
If
ABSOLUTELY SURE
the connector at the
2
nd
position from the
LEFT, “MON-NTC”, is unplugged AND [
Temperature
] is STILL more than 30.0° C proceed to
page
, procedure number
T- 1.0.131 [TEMPERATURE] WINDOW LESS THAN 30.0° C
a) Return NTC #3’s connector to the
2
nd
distribution board
position from the LEFT, “MON-NTC”.
b) If NTC #3 was plugged in correctly
[Temperature]
RETURNS to between 35.1 and 38.9° C!
(page 225).
T- 1.0.132 VERIFY CONDUCTIVITY
a)
ENSURE the shunt door is closed!
b) Based on the [
Conductivity
] window, THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) STABLE* between 13.0 and 14.5 mS AND the window is pale yellow/white:
(page 226).
* STABLE = NOT changing more than 0.2 mS per minute
2) IF (and ONLY if) less than 13.0 OR more than 14.5 mS:
Proceed to
page 317
SECTION 5 – CONDUCTIVITY PROBLEMS
3)
IF STABLE* between 13.0 and 14.5 mS AND the window IS RED:
See parts a THROUGH c
below:
a) At the bottom of the screen, select the
‘Dialysate’ tab (Figure right).
b) Figue right, if necessary, adjust the Alarm
Limits until ‘Actual’ Conductivity is CENTERED
between them.
c) Press ‘CONFIRM’! Is the [
Conductivity
]
window white?
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...