2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 118
F- 15.2.2 MOTOR CONTINUES TO ROTATE
Recheck Loading Pressure (Rinse Port gauge). TWO (2) possible scenarios below:
1)
IF (and ONLY if) REMAINING LESS THAN
15 psi:
2)
IF peaking to at least 15 psi:
NOTING how
many turns, turn Loading Pressure Valve #65’s
nut (Figure right) CLOCKWISE (inward) until if
pressure reaches a PEAK of between 23 and
25 psi. TWO (2) possible scenarios i) or ii)
below:
i)
IF (and ONLY if) a high of between 23
and 25 psi CANNOT be achieved:
Return Valve #65’s nut to its ORIGINAL
location then see procedure number
(page 119).
ii)
IF between 23 and 25 psi CAN be
achieved:
See parts a THROUGH d below:
a) Figure right, if threads are visible under
Valve #65’s nut see part b. If no threads
are visible either the wrong spring is
installed* OR Valve #65 is bad.
(page 19)
b) From the Home screen, select the
[Dialysate Flow] window.
c) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
d) Call debug screen 6.
e) WITHOUT LOOKING AWAY, watch
BC Switch
(middle column) for five (5)
minutes OR until it EVER = 897, or more even if only once?
Yes
BC Switch
goes to 897 or more! Return to
page 30
, procedure number
No
BC Switch
NEVER = 897 or more! See parts a AND b below:
a) Turn the Heater Switch on!
b) Check Deaeration Pressure
per the Preventative Maintenance Procedures
booklet.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...