2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 119
F- 15.2.3 LOW LOADING PRESSURE / ISOLATE POSSIBLE AIR LOCK
a)
Obtain a 1000 ml graduated cylinder AND a
bucket!
b) Figure right, unplug the DERATION PUMP’s,
connector from distribution board position “P20-Degas-
P”!
c) Figure right, remove the
clear tubing
from the
Deaeration Pump’s INPUT nozzle.
Keep the
tubing below the pump to encourage
gravity flow!
d) The tubing is from hydrochamber D. If ANY FLOW from it
direct it into the bucket for TWO (2) minutes BEFORE
continuing to part e.
e) Now direct the tubing into the cylinder for ONE (1) minute.
MORE THAN one hundred thirty (130) ml collected?
Yes
More than 130 ml per minute! See procedure
number
(page 119).
No Less than 130 ml per minute! The Deaeration Pump may be air locked! Leaving the tubing off
AND the pump unplugged, proceed to
page 122
, procedure number
F- 15.2.4 MORE THAN 130 ML PER MINUTE COLLECTED I.E NO AIR LOCK
a)
Reattach the Deaeration Pump’s tubing!
b) Return the Deaeration Pump’s connector to distribution board position P20, “Degas-P”.
c)
ENSURE the DEAERATION MOTOR is running!
(page 120).