ELYRA 7
Left Tool Area and Hardware Control Tools
ZEISS
03/2019 V_02
000000-2262-999
99
5.2.10.4
Display Options
You can set display options by opening the
Display
expansion field (see Fig. 143).
If the
Auto copy from last acquired
box is
checked, the settings from the last acquired image
will be taken.
By checking the
Range indicator
box the image
will be displayed with the range indicator look up
table.
5.2.11
Acquisition Parameter – Focus
If the
Focus
tool tab is expanded the
Focus
control
tool becomes available (Fig. 144).
The
Z-Position
display box shows the current
z-position.
The number can be edited by typing or using the
arrowheads. The nosepiece or stage will
automatically update the new position.
By pressing the
Stop
button, you can stop the
current movement. The
Step Size (µm)
display
box shows the currently used step size for focus
movement. The value can be edited by typing or
using the arrowheads.
The
Home
button sets the z-position to the
predefined zero value.
Pressing
Load
will lower the nose piece (z-position)
for objective change.
Work
brings the z-position back to the actual used
position if the focus was zeroed or in
Load
position.
Home and Load positions can also be accessed
from the z-position display by pressing the
H
or
L
buttons.
The z-position can be altered by grapping and sliding the
plane
or by using the arrow slider. The
plane
shows the actual z-position.
Pressing
Manually
in the
Zero position
field will set the current z-position to 0.
If a
Z-Piezo stage
insert is in use and active then pressing
Leveling
in the
Piezo field
will center the
Piezo stage position in regard to the microscope focus drive.
The
Piezo Position (
µ
m)
input box allows the position of the Piezo to be altered.
The
Step Size (µm)
display box shows the currently used step size for focus movement of the
Z-Piezo
stage
.
Fig. 143
Display expansion field
Fig. 144
Focus tool