5.2.20.4
End
Expand the
End
field by clicking the
triangle (Fig. 226).
In the
Mode
drop down menu
Fig. 227) you can select the way the
acquisition is triggered. You have the
choice between Manual, Trigger and
Time.
If
Manual
is selected the acquisition of
a time series ends using the
Stop
button
or after the defined number of
cycles.
If
Trigger
is selected the
Trigger In
drop down menu becomes available
(Fig. 228) from which up to 4 triggers
can be selected. The system will
interrupt the acquisition once the
trigger in signal is received.
If
Time
is selected the
hh
,
min
and
sec
input boxes become available
(Fig. 229), in which the hours, minutes
and seconds can be edited. The system
interrupts the acquisition after the
defined computer time has been
elapsed.
It is possible to send a
Trigger Out
signal with the end of the times series.
The
Trigger Out
drop down menu
allows defining up to 4 trigger out
signals.
If the entered number of cycles is processed without a set trigger impulse been given to end the
procedure, the Time Series is finished.
If a trigger signal to end the scan arrives before the cycles have been processed, the Time Series will
only be interrupted.
Waiting for Trigger
will be displayed in the status line. The Time Series can
now be continued via a new trigger signal or ended via
Stop
.
A Time Series is finished when the end
Time
is reached.
If the entered number of
Cycles
has been processed, the Time Series is finished before the end
Time. If the number of cycles has not yet been processed until the end Time, the Time Series is also
finished. The end time for the Time Series can be changed online. Bleach times will be added to the
total time for image acquisition.
If a time series is interrupted before its programmed end, the programmed number of images will
be taken over in the database. However, only those images are stored which were created before
interruption of the time series. This is due to the fact that the original image parameters are to be
taken over via the
Reuse
function.
It is possible to stop and restart a time series via TriggerIN and have the data concatenated in one
file by alternating between the Start and End function. For example starting the time series with
Fig. 226
End field
Fig. 227
End Mode drop down menu
Fig. 228
End Mode drop down menu – Trigger selected
Fig. 229
End Mode drop down menu – Time selected