ELYRA 7
280
000000-2262-999
03/2019 V_02
Ticking the
Exchange x and y
check box will exchange x for y coordinates and vice versa. The original
data set is not changed. The
Fast
image acquisition works differently than conventional confocal point
scanning. One difference is the 90° rotated scan direction. In order to match
Fast
images with other
acquisition mode and the driving behavior of scanning stages, the
Exchange x and y
checkbox always
needs to be checked.
Ignore file sets
provides the possibility to open all files, including the master file, as a separate file.
Master files are generated when images are saved in individual files along time point, stage position, tile,
Z stack position and/or channel. This is defined within the Autosave Tool. Per default the option is not
active and such subfiles are opened as one file if the master file (the first file set of all) is opened. Only the
subsequent files can be opened individually unless this function is active.
Note: When a file set is opened via the masterfile and modified (i.e. cut region, subset etc.) and then
saved again, the new data set will be saved only in one file and not in file subsets like the original data.
Airyscan Processing Baseline Shift
- not available.
5.4.4.7
System Options – Streaming
Press the
Streaming
tab to open the
Streaming
panel (Fig. 476).
To increase the performance of the computers
when opening and saving files, the
Use
Streaming to open CZI files
checkbox needs to
be checked.
This allows the system to access files faster from the streaming location.
Streaming in general is managed in the
Streaming and Auto Save
tool (see section 5.2.23).
5.4.5
Options tool group – Experiment
Designer
During the acquisition of multiblock experiments
the currently acquired images of a block are
displayed in the image container.
This leads to a constant switching between the images during the acquisition. If this is not wanted, it can
be deactivated by clearing the
Automatic Document switch
check box in this tool window (Fig. 477).
5.4.6
Options tool group – Service
Only for service use.
Fig. 476
System Options – Streaming panel
Fig. 477
Maintain, Experiment Designer
Fig. 478
Options, Service