109
Info on Status
The
Status:
line indicates the current status of the
Definite focus. It comprises the following messages:
−
Device not initialized
(in red): Definite Focus
is switched off or experienced hard- or
firmware problems during initialization.
−
Error
(in red): occurs when the Definite Focus
has no sufficient signal or when there is an
error on the hard- or firmware (Fig. 167).
−
Reflex Finder running
(in blinking green):
whenever the
Find Surface
button was
pressed and the system is actively looking for
the glass reflex.
−
Reflex found
(in green): whenever after
pressing the
Find Surface
button a glass reflex
was found.
−
Reflex not found
(in grey): whenever after
pressing the
Find Surface
button the glass
reflex was not found (due to missing signal).
−
Stabilizing
(in blinking green): whenever the
Periodic
button was pressed for continuous
stabilization
−
Stabilizing
(in green): whenever the
Periodic
button was pressed for interval stabilization.
−
Position stabilized
(in green): whenever the
Recall Focus
button was pressed and the Definite
Focus found a sufficient strong signal to stabilize on.
−
Position not stabilized (in grey)
: whenever the
Recall Focus
button was pressed and the
Definite Focus did not find a sufficient strong signal to stabilize on.
−
Determine position
(in blinking green): whenever the
Recall Focus
button was pressed and the
Definite Focus screens for a sufficient strong signal to stabilize on.
The
- Error -
message will be listed in the Error log display. The Definite Focus device shows in its
display also
Error
followed by a number that can be looked up in the Definite Focus manual.
Fig. 166
Warning information at ZEN Software startup if the system
detects differences in the ZEN and Firmware listed
objectives
Fig. 167
Definite Focus tool: "Status: " line