ZEISS
Left Tool Area and Hardware Control Tools
ELYRA 7
262
000000-2262-999
03/2019 V_02
5.3.24.3
Channel Shift
Channel shift
is used to produce a congruent
image with relation to the pixels of the various
channels (Fig. 439).
This pixel correction function is particularly
important in UV applications.
This is a manipulation of your data that has
to be considered very carefully when co-
localization analysis is part of your
experiment.
The
Select
button selects the active image in the
Image Display
as the Input Image.
The image has to be a multi-channel image.
•
Select the channels you want to shift by ticking
the Ch1-T1 or Ch2-T2 check box:
•
Use the X and Y sliders or spin boxes to define
the shift in pixels to select the pixel shift in the
horizontal and vertical direction. Clicking the
Zero next to the spin-boxes resets the shift to
the original position.
A preview is automatically generated at the
bottom of the
Channel Shift
tool in the
Processing
tab of the Left Tool Area At the same
time, a large preview is shown in the
Preview
view tab in the Image display in the Center Screen
Area.
•
Click on the
Apply
button to generate a new
image with the set pixel shift.
Fig. 439
Processing – Adjust with Channel shift