ZEISS
Left Tool Area and Hardware Control Tools
ELYRA 7
114
000000-2262-999
03/2019 V_02
•
Go on with this procedure for marker two and three. Use coordinates for calculating the distances
between the markers.
Mounting of cover glasses or TEM grids
•
Place the specimen (sample carrier) in the
suitable sample holder and fix it:
Life Science Cover Glass 22x22:
−
Remove the clamping frame (Fig. 177/
1
)
using tweezers.
−
Insert the cover glass (Fig. 177/
2
) in the
sample holder (Fig. 177/
3
).
−
Slide in the clamping frame into the sample
holder until the clamps are clicking into place
(Fig. 177/
4
).
Life Science for TEM Grids:
−
Lift the spring (Fig. 177/
5
) of the appropriate
position and turn it side wards.
−
Insert the TEM grid (Fig. 177/
6
) into the
provided holding spot of the holder
(Fig. 177/
7
) and fix it with the spring.
Fig. 177
Mounting of cover glasses or
TEM grids