ZEISS
Center Screen Area / Image Containers - Display and Image Analysis
ELYRA 7
350
000000-2262-999
03/2019 V_02
6.16
Unmixing View
In the
Unmixing View
, Lambda or channel data
sets can be analyzed and spectrally unmixed (for
more information see as well section
). In particular it allows to:
−
display the spectra corresponding to user-
defined ROIs (mean ROI intensity over
Lambda),
−
use all types of multi-channel images as
input,
−
show the intensity values in table form and
copy table to clipboard or save as text file,
−
generate unmixed multi-channel images.
The available tools and the functionality of the four
general View Option control blocks
Dimensions
,
Display
,
and
Graphics
is the same as in the
Lambda Coded
View (see the previous section on
this view type). The two additional view-specific
Unmixing View Option
and
Unmixing Extract
View Option
control blocks are shown in Fig. 563
and Fig. 564.
Any changes done with this View Option control
blocks are effective immediately.
To access the Unmixing analysis view type, load or acquire a Lambda stack data set or a multichannel
image and click on the
Unmixing
tab.
Detector calibration is a service task. This calibration allows that reference spectra can be acquired
at different detector gain settings than the multi-labeled image acquired for unmixing!
Fig. 563
Unmixing View Option control block
Fig. 564
Unmixing - Extract - View Option
control block