Expand To Maximum Volume
: This checkbox is activated as default. In a multiview experiment not all
views might include all the information. When
Expand To Maximum Volume
is activated, the volume
of the resulting dataset after registration and fusion is determined and additional black pixel (x,y) and z-
slices are added to the views during registration. This prevents loss of data for later fusion.
Scaling x=y=z
: The dimensions are downsized to the smallest to have the same size.
Individual Registration For Every Time point
: If the multiview experiment has been performed over
time, this option becomes available.
−
If
activated
,
each time point generates a transformation matrix and each time point is registered
individually. As a result, the time necessary for registration is multiplied by the number of time
points and can rise substantially. However, if the motorized movement of the sample by the system
has changed (e.g. due to thermic effects on the system: not fully warmed up at start, incubation,
air conditioning of the room), the Individual Registration For Every Time point option becomes
necessary.
−
If
not activated
, one transformation matrix is generated for all the following time points based
upon the first time point. As a result, processing for registration has only to be done once which
reduces processing time.
Fig. 416
Dual Side Fusion – Fusion Subset in X, Preview and
Image control area