ELYRA 7
Left Tool Area and Hardware Control Tools
ZEISS
03/2019 V_02
000000-2262-999
145
Select the carrier configuration from the
Configurations
list by highlighting it or
search for it by typing its name in the
Search
field. The
Info
window pops up Fig. 237).
Press
Yes
to delete the configuration,
No
to
keep the configuration. To save a carrier
configuration press the
Save
button
(symbolized by the
disc
icon).
This opens the
Save current carrier
configuration as
panel (Fig. 238). Type in
an appropriate name in the
Configuration
field and press
Ok
to save the configuration
in the data base or press
Cancel
to abort the
process.
To assign the actual position of the stage
with a position in the sample carrier, move
to the center of a well in the actual plate
under the microscope then click
Calibrate
followed by clicking the field in the sample
carrier representing this well of the plate.
This assignment will be kept until changed
repeating this procedure and that well will
turn blue.
Pressing the
?
button (Fig. 239) will display a
help text how to calibrate the position of the
sample carrier.
If you click on any other than the blue
highlighted well, the stage will move to this
well based on the distances defined in the
carrier properties. This will be than the blue
highlighted one.
By pressing
Select
you can highlight those
wells of the sample carrier that should be
sequentially used by just clicking on them
(Fig. 240). All selected wells will turn blue. By
clicking on a selected blue highlighted well,
it will be deselected. By pressing
Select all
,
all wells will be selected for imaging. By
pressing
Clear all
, all wells will be
deselected. As long as Select stays pressed,
the wells will be sequentially approached
starting at the let upper position to the right
lower one (A1, A2…;B1, B2,…).
Fig. 237
Delete information window
Fig. 238
Save current carrier configuration as window
Fig. 239
Calibration of sample carrier
Fig. 240
Selection of wells in a sample carrier