Left Tool Area and Hardware Control Tools
ELYRA 7
100
000000-2262-999
03/2019 V_02
Use of an optional high resolution focusing device (Piezo objective focus or Z-Piezo stage
insert)
The
Piezo Position
is indicated as a numerical value. This value and with it the focus position can be
changed. A change of this value updates the value of the microscope focus accordingly.
The
Step Size
, by which the Piezo focus moves when clicking on the arrows, can be set between
0.005 µm and 10 µm.
As soon as the focus position of the microscope is changed (via hand wheel or software), the piezo
objective focus is automatically levelled.
A click on the
Leveling
button moves the Piezo objective focus or Z-Piezo stage insert into the center
position of its travel range and the focus position of the microscope is reset to compensate for the
leveling. Therefore, the same Z-level remains visible (the current position is not set to zero).
The motor focus of the stand is operated in the same way via the relevant buttons. Moving into the
Work
or
Load
position is always performed via the motor focus and not via the Piezo objective focus.
Please see printed manual CHAPTER 4: ANNEX for further information on the Piezo objective focus
and Z-Piezo stage insert.