4-12
Operator’s Manual – OPTI CCA-TS2
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-23 QC Results
•
The QC sample is aspirated into the cassette,
and then measurement starts (Fig. 4-22).
At this time the status light begins flashing
green indicating that the cover should not be
opened.
•
Upon completion of the measurement, the
results are displayed (Fig. 4-23).
•
The OPTI CCA-TS2 Analyzer will indicate
whether the values are within or outside the
programmed ranges with a
<Pass/Fail>
display next to the parameter label.
•
Results obtained are applicable to the sensor
cassette type being used for patient sample
(E-Ca Type shown for reference).
18. Press
(Fig. 4-23) to accept or reject
results.
•
In the subsequent screen (Fig. 4-24),
you can press
<Accept>
if results are
acceptable, and the results will be stored in
the Control Database.
•
Select
<Reject>
to reject the results.
Rejected results will not be stored in the
Control Database.
•
Select
<Review>
to view the results again.
NOTE: In either case, the results will be printed
when the data input is complete. Please
follow the regulatory guidelines of your
hospital for documenting corrective action,
if results are rejected.
NOTE: Data will be exported using the configured
export method (see Section 3.2.4.5) when
the results are printed.
NOTE: The automatic printout feature may be
turned off in setup (See Section 3.2.4.1).
Additional information on printing reports
and exporting data can be found in Chapter
6, Data Manager.
•
For troubleshooting, refer to Chapter 8.
19. When prompted, open the sample chamber
cover and remove the cassette.
•
If other levels of controls are to be run,
repeat the procedure.
Fig. 4-22 QC Measurement
Fig. 4-24 Accept QC Results