Operator’s Manual – OPTI CCA-TS2
PCO2-3
ANALYTES
P
CO
2
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pCO
2
sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %CO
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
72
Slope
0.98
Offset
0.45
R
2
0.999
Sy|x
0.975
0
50
100
150
200
250
0
50
100
150
200
250
PC
O
2
-T
S2
PCO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(0.45 + 0.98x)