Operator’s Manual – OPTI CCA-TS2
Ca-5
ANALYTES
CALCIUM
Correlation to Other Methods
4
OPTI CCA-TS vs other Electrolyte Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for electrolyte analyses were analyzed by both traditional
and non-traditional operators of measurement equipment in hospital laboratories. The sample was
analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for
these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
0.8732
-0.064
0.8392
0.07
0.7 to 1.3
103
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.