Operator’s Manual – OPTI CCA-TS2
pH-5
ANALYTES
pH
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
126
Range
6.927 to 7.705
Slope
0.97
Offset
0.18
R
2
0.998
Sy|x
0.008
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests,
2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Peterson JI, et.al.,
A Fiber Optic pH Probe for Physiological Use
, Anal.Chem. 53,p.864, 1980
.
3. Wolfbeis OS, Offenbacher H,
Fluorescence Sensor for Monitoring Ionic Strength and Physiological
pH Values
, Sensors and Actuators 9, p.85, 1986.
4. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)