Operator’s Manual – OPTI CCA-TS2
8-29
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.18 Checking the Touch Screen
The purpose of this test is to check the proper
operation of the touch screen.
In the main menu, press
<System Manager ->
Diagnostics>
.
•
Select the
<Tests>
tab.
•
Press
<Touch>
to start the test (Fig. 8-30).
•
Touch the screen and a dot should appear under
the touched location (Fig. 8-31).
•
If not, press
<Calibrate>
to perform a touch
calibration.
•
Using a finger, stylus or pointed object (e.g.
syringe adapter), touch the center of the
calibration mark as it moves around the screen
(Fig. 8-32).
NOTE: Do not use sharp objects, since they may
damage the screen.
•
When finished press
.
•
Press
to return to the
<System>
screen
or
to return to the main menu.
Fig. 8-30 Select Touch Test
Fig. 8-31 Perform Touch Test
Fig. 8-32 Touch Calibration
Содержание CCA-TS2
Страница 1: ...OPTI CCA TS2 Analyzer Operator s Manual ...