PCO2-2
Operator’s Manual – OPTI CCA-TS2
P
CO
2
ANALYTES
Reproducibility
Controls
Within-run precision (S
wr
) and Total Precision (S
T
), were determined from 2 runs per day with 2
replicates per run over a period of 20 days following the CLSI guideline EP5-A2. Typical results for 3
different control levels are shown below:
PCO
2
(mmHg)
OPTI Check Level 1
OPTI Check Level 2
OPTI Check Level 3
Days Run
20
20
20
Total Average
74.5
45.0
24.8
Within Run St. Dev. (S
wr
)
0.8
0.3
0.3
Within Run % CV
1.1%
0.7%
1.1%
Total Precision St. Dev. (S
T
)
0.9
0.5
0.4
Total % CV
1.3%
1.0%
1.5%
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on the OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
PCO
2
(mmHg)
Whole Blood
Level 1
Level 2
Level 3
Average
21.4
46.2
91.6
St. Dev
0.3
0.4
1.7
%CV
1.60%
0.90%
1.82%
n
10
10
10