PCO2-B-2
Operator’s Manual – OPTI CCA-TS2
P
CO
2
(Dry Sensor)
ANALYTES
Interferences
The following exogenous interferents were quantified in tonometered whole blood samples
spiked with a number of endogenous and exogenous chemicals and tested for interference
following the CLSI guideline EP7-A2, showing interferences to dyes which typically have short
half-lives within the body before being metabolized by the liver. This testing was done on the
OPTI CCA-TS platform, with a standard OPTI style pCO
2
sensor. No changes were made to the
sensor or the measurement principle, so the interference of the OPTI CCA-TS2 is equivalent to
the OPTI CCA-TS.
Chemical
Interferent
Concentration
PCO
2
Level
Interference
Acetaminophen
1.66 mM
83 mmHg
NO
17 mmHg
NO
Acetylsalicylic acid
3.33 mM
83 mmHg
NO
17 mmHg
NO
Ascorbic acid
0.23 mM
83 mmHg
NO
17 mmHg
NO
B-Hydroxybutyric acid
16.03 mM
83 mmHg
NO
17 mmHg
NO
Bilirubin
0.26 mM
83 mmHg
NO
17 mmHg
NO
Cardiogreen
0.0065 mM
83 mmHg
NO
17 mmHg
NO
Cystein
6.41 mM
83 mmHg
NO
17 mmHg
NO
Ethanol
86.8 mM
83 mmHg
NO
17 mmHg
NO
Evans blue
0.0104 mM
83 mmHg
-24.68 mmHg
17 mmHg
NO
Glycolic acid
10 mM
83 mmHg
NO
17 mmHg
NO
Halothane
0.759 mM
83 mmHg
NO
17 mmHg
NO
Ibuprofen
2.43 mM
83 mmHg
NO
17 mmHg
NO
Intralipid
1%
83 mmHg
NO
17 mmHg
NO
Methylene Blue
0.125 mM
83 mmHg
NO
17 mmHg
NO
Sodium Chloride
20 mM
83 mmHg
NO
17 mmHg
NO