Operator’s Manual – OPTI CCA-TS2
7-7
7 MAINTENANCE
Fig. 7-17 Push on Pump Roller
5. Press the pump cartridge roller down until it
firmly sits on the shaft of the pump motor
(Fig. 7-17).
6. Perform a
<Pump Test>
(see section 8.2.16)
to ensure correct operation. Make sure the
pump rotates smoothly without excessive noise.
In addition, run one sample in control mode.
Make sure the control measurement passes
without errors.
The OPTI CCA-TS2 has a function that allows you
to select maintenance reminder options which will
alert you when the next pump replacement is due.
Any maintenance actions that you perform through
the maintenance reminders will be captured in the
maintenance log of the analyzer.
1. From the
<Ready>
display, select
<System Manager>
and
<Maintenance>
.
2. Enter User ID if enabled
3. In the
<Maintenance>
menu (Fig. 7-18),
press
<Setup>
.
4. In the
<Setup>
screen (Fig. 7-19), you
can select the option
<Replace Pump
Reminder>
.
The analyzer will then remind you when the
next pump replacement is due.
5. After you replace the peristaltic pump, go to the
<Maintenance>
screen (Fig. 7-18) and press
next to
<Pump Replacement>
.
The date of the last pump replacement will be
displayed for future reference.
Fig. 7-18 Maintenance Setup
Fig. 7-19 Replace Pump