Operator’s Manual – OPTI CCA-TS2
8-19
8 DIAGNOSTICS AND TROUBLESHOOTING
8.2.2 Checking System Temperatures
In the main menu, press
<System Manager ->
Diagnostics>
.
The
<Temperature Readings>
option lets you
check the various system temperatures:
<Top Plate>
,
<Bottom Plate>
and
<Ambient>
(Fig. 8-3).
NOTE: If top or bottom plate temperatures are
out of range, the temperature display will
change to red.
•
Press
to return to the
<System>
screen
or
to return to the main menu.
Fig. 8-3 Temperature
8.2.3 Checking Gas Pressure
Fig. 8-4 Gas Pressure
The
<Gas Level>
option allows you to check the
percent remaining of the gas bottle (Fig. 8-4).
With a new gas bottle in place, the pressure should
be approx. 99%, with the bottle removed, the
pressure should be 00%.
In the main menu, press
<System Manager>
and
<Diagnostics>
.
•
To install a new gas bottle, press
.
•
Scan the gas bottle bar code located on the insert
sheet to install a new gas bottle (see Section
7.5.1 “Changing Gas Bottle”) .
•
Press
to return to the
<System>
screen
or
to return to the main menu.