Operator’s Manual – OPTI CCA-TS2
THB/SO2-3
ANALYTES
tHb/SO
2
Interferences
The following exogenous interferents were quantified in tonometered whole blood, showing interferences
to dyes which typically have short half-lives within the body before being metabolized by the liver.
This testing was done on the OPTI CCA platform, with a standard OPTI style cassette. No changes were
made to the sensor or the measurement principle for the tHb and SO
2
measurement, so the interference of
the OPTI CCA-TS2 is equivalent to the OPTI CCA.
EXOGENOUS
Substance
Amount
tHb change (g/dL)
SO
2
change (%)
Cardio (indocyanine) green
0.5 mg/dL
+4.7
+4%
Evan’s Blue
5.0mg/dL
<1.0
-17%
Methylene blue
25 mg/dL
+3.0
-37%
ENDOGENOUS
Substance
Amount
tHb change (g/dL)
SO
2
change (%)
Carboxyhemoglobin
10%
-2.0
<2%
Carboxyhemoglobin
20%
-3.3
<2%
Methemoglobin
13%
+1.7
-7%
Rapidly sedimenting blood samples should be mixed thoroughly and immediately aspirated into the
OPTI cassette, as described in Section 5.1.7 “Handling and Storage of Samples”, to ensure accurate tHb
measurements. If allowed to sediment, the blood sample’s reported tHb may be falsely high or low.
Fetal hemoglobin taken from cord blood extracts was tested and showed no interference to the tHb and
S
O
2
measurement.
Only clear, uncolored quality control materials, such as OPTI CHECK or OPTI CHECK PLUS brand
aqueous controls should be used with the OPTI CCA-TS2 system. Colored materials, including
proficiency testing materials, may interfere with the tHb and SO
2
measurement, or fail to be properly
aspirated.