Operator’s Manual – OPTI CCA-TS2
pH-B-5
ANALYTES
pH (Dry Sensor)
Correlation to Other Methods
3
OPTI CCA-TS vs other pH Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed
on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for these
analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
OPTI R
0.96
0.27
0.984
0.019
6.78 to 7.54
147
Analyzer A
1.03
0.20
0.968
0.015
7.091 to 7.538
111
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
OPTI CCA-TS with B-Lac Cassette vs other pH Instruments on whole blood
(in-house testing)
Whole blood samples from multiple donors were tonometered with different %CO
2
gas mixtures to
generate a wide range of pH values. The blood samples were analyzed in parallel on the B-Lac cassette
and other laboratory instruments.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer B
1.03
-0.21
0.996
0.016 6.578 to 7.766
174
Analyzer C
1.03
-0.19
0.996
0.015 6.582 to 7.701
174
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.