Operator’s Manual – OPTI CCA-TS2
7-3
7 MAINTENANCE
4. Gently clean the optics window and the inside
top cover of the sample chamber with a soft lint
free cloth. Press
(Fig. 7-6).
5. At the prompt, open the SMC cover by pressing
down on the red latch (Fig. 7-7).
6. Gently wipe both sides of the tHb-Calibrator
Cassette with a clean dry cloth and examine it to
ensure it is clean. Insert it into the chamber and
press down to properly seat the cassette
(Fig. 7-8).
Fig. 7-7 Open Cover
Fig. 7-8 Wipe and Insert Cassette
Fig. 7-6 Clean Optics