Operator’s Manual – OPTI CCA-TS2
9-15
9 OPERATING PRINCIPLES
Calibration Gas, BP7162
Use:
For calibration of pH,
P
CO
2
and
P
O
2
in the OPTI CCA-TS2 Analyzer.
Contents:
Each disposable, low-pressure cylinder contains 0.35 liters of gas
at 28 psi at 21 °C.
Composition:
Oxygen
14.0 ± 0.02%
Carbon Dioxide
6.0 ± 0.02%
Nitrogen
balance
Storage:
Refer to package labeling.
Hb Calibrator Cassette, BP7653
Use:
For quarterly calibration of the OPTI CCA-TS2 Analyzer.
Contents:
Each package contains one reusable calibrator cassette.
Composition:
Stabilized optode sensors
Storage:
Refer to package labeling.
Stability:
Expiration date and lot number are printed on each package label and
encoded on the attached bar code label.
Precautions
Use of calibration solutions, calibration gas, or optodes not manufactured by OPTI Medical
Systems could void the warranty.
Once used, the sample cassette holds human body fluids which may be potentially infectious;
handle with appropriate care to avoid skin contact or ingestion.
For
in-vitro
diagnostic use.
For professional use only.
Bibliography
1.
Guilbault GG, Ed.,
Practical Fluorescence
, 2nd Edition, Marcel Dekker, 1990
2.
Tietz; Burtis C, et al (Eds.),
Textbook of Clinical Chemistry and Molecular Diagnostics
, 4
th
Ed.,
(Elsevier Saunders, 2006) pps. 2252-2302.
3. Mahoney JJ, Wong RJ, Van Kessel AL,
Reduced Bovine Hemoglobin Solution Evaluated for Use as a
Blood Gas Quality Control Material
. Clin.Chem.39/5,874-79 (1993)
4. J.B.Henry,
Clinical Diagnosis and Management by Laboratory Methods
, 19th Ed., 1996, p.590,777
5.
Clinical and Laboratory Standards Institute (CLSI).
Interference Testing in Clinical Chemistry;
Approved Guideline - 2nd Edition
. CLSI document EP7-A2. CLSI, Wayne, PA, 2005