Operator’s Manual – OPTI CCA-TS2
PO2-B-5
ANALYTES
P
O
2
(Dry Sensor)
Correlation to Other Methods
4
OPTI CCA-TS vs other Blood Gas Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of blood gas equipment in hospital laboratories. The blood was analyzed
on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for these
analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
OPTI R
1.04
-2.76
0.968
8.72
27 to 288
148
Analyzer A
0.97
3.73
0.992
5.24
27.0 to 423.8
110
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.
OPTI CCA-TS with B-Lac Cassette vs other Blood Gas Instruments on whole blood
(in-house testing)
Whole blood samples from multiple donors were tonometered with different %O
2
gas mixtures to
generate a wide range of pO
2
values. The blood samples were analyzed in parallel on the B-Lac cassette
and other laboratory instruments.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
OPTI CCA
0.94
4.84
0.998
9.77 19.9 to 642.8 161
Analyzer B
0.95
6.32
0.992
18.25 17.0 to 635.7 161
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.