8-6
Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
8.1.2 System Error Messages
<Maximum Users Exceeded>
The number of user IDs stored in memory has
reached 300.
•
Press
to continue.
•
Delete unused user IDs from memory
(See Section 3.2.3.1.2).
<ERROR! - Bad Sensors>
Two or more measured parameter sensors are bad.
•
Press
, discard the cassette and repeat
the test with a new cassette.
<ERROR! - Possible Clot>
A sample error has occurred. This may be due to a
clot or blockage preventing sample aspiration.
•
Press
and discard cassette.
<ERROR! - B-Lac Setup failed>
This error message appears when the reproducibility
of the OPTI Check controls during the lactate setup
procedure is out of range.
•
Repeat lactate setup procedure. Make sure
to aspirate the OPTI Check directly from the
ampoule.
•
Repeat the lactate setup using a different lactate
cassette lot.
•
Call Technical Support.
<ERROR! - Wrong/Different Lot>
An incorrect cassette type or lot was scanned in
during B-Lac setup.
•
All cassettes used during lactate setup must be
the same lot number.
•
Make sure the cassette type is B-Lac.