BUN-4
Operator’s Manual – OPTI CCA-TS2
BUN/UREA
ANALYTES
Direct linearity for the OPTI CCA-TS2 measurement has been established against reference materials or
methods. The linearity of the BUN (urea) sensor on the OPTI CCA-TS2 has been established versus the
same sensor on the OPTI CCA-TS using whole blood samples tonometered with mid level mixtures and
spiked or diluted to establish the correlation.
Parameter and Specification
Value
Number of Samples
63
Slope
1.00
Offset
0.08
R
2
0.998
Sy|x
1.703
0
20
40
60
80
100
120
140
160
0
20
40
60
80
100
120
140
160
B
U
N
-
TS2
BUN - TS
Scatter Plot with Fit
Identity
Linear fit
(0.08 + 1.00x)