Operator’s Manual – OPTI CCA-TS2
PO2-5
ANALYTES
P
O
2
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
144
Range
13.5 to 639.9
Slope
0.98
Offset
-1.82
R
2
0.999
Sy|x
3.931
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Kautsky H,
Quenching of Luminescence by Oxygen
, Transactions Faraday Society 35, p.216, 1939
3. CDI,
3M Healthcare System 200 Extracorporeal Blood Gas Monitor
. See, for example, Lubbers DW,
Gehrich J, Opitz N,
Fiber Optics Coupled Flourescence Sensors for Continuous Monitoring of Blood
Gases in the Extracorporeal Circuit
, Life Supports Systems 4, p.94, 1986.
4. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).