PCO2-B-6
Operator’s Manual – OPTI CCA-TS2
P
CO
2
(Dry Sensor)
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
63
Range
14.7 to 87.9
Slope
1.00
Offset
-1.13
R
2
0.995
Sy|x
1.523
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
10
20
30
40
50
60
70
80
90
10
20
30
40
50
60
70
80
90
D
ry
PC
O
2
-T
S2
Dry PCO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(-1.13 + 1.00x)