8-2
Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
<WARNING! - No Sample Detected>
The sensors did not detect any sample.
•
Make sure the sample is properly attached and
not clotted and does not contain air bubbles.
•
Wait for the system to recalibrate.
•
Remix the sample carefully.
•
If the system does not detect the sample after
retrying, press
to notify the system
that the sample is reattached and reaspirate
sample.
<WARNING! - Unstable pH>
The displayed measured parameter is unstable.
NOTE: This message is a warning. The analyzer
will, however, display a result for the
parameter concerned.
•
For lactate cassettes, you have the option
of continuing the measurement by pressing
or stopping by pressing
.
•
For all other cassettes, press
to
continue.
•
Once the measurement is complete, remove
cassette and check for aspirated bubbles.
•
If bubbles are present over a sensor, do not
report that parameter.
<WARNING! - Bad Sensor pH>
The displayed sensor is defective.
•
For lactate cassettes, you have the option
of continuing the measurement by pressing
or stopping by pressing
.
If you continue, no results will be provided
for the defective sensor or any calculated
result, which
utilizes this measurement in its
calculation.
8.1.1 System Warning Messages