Operator’s Manual – OPTI CCA-TS2
7-1
7 MAINTENANCE
7 MAINTENANCE
7.1 Daily Maintenance
No daily maintenance is required for the OPTI
™
CCA-TS2 system.
7.2 Weekly Maintenance
Once a week, the Sample Measurement Chamber (SMC) must be cleaned.
Open the top cover and clean the optics surface as well as the underside of the SMC
cover with a lint-free cloth, dampened with a dilute alcohol or ammonia-based
cleaner as needed. Be sure to remove all blood residue with a 10:1 diluted bleach
solution. A cotton swab may be used for cleaning the smaller parts of the SMC.
Fig. 7-1 Maintenance Setup
Fig. 7-2 Select Cleaning
The OPTI CCA-TS2 has a function that allows you
to select maintenance reminder options which will
alert you when analyzer cleaning is due.
Any maintenance actions that you perform through
the maintenance reminders will be captured in the
maintenance log of the analyzer.
1. From the
<Ready>
display, select
<System Manager>
and
<Maintenance>
.
2. Enter User ID if enabled.
3. In the
<Maintenance>
menu (Fig. 7-1),
press
<Setup>
.
4. In the
<Setup>
screen (Fig. 7-2), you can
select the option
<Cleaning Reminder>
.
Select
<Weekly>
for weekly maintenance and
<Monthly>
cleaning if the analyzer is not used
on a weekly basis.
The analyzer will then remind you when the
next analyzer cleaning is due.
5. After you perform the analyzer cleaning
procedure, go to the
<Maintenance>
screen (Fig. 7-1) and press
next to
<Cleaning>
.
The date of the last cleaning will be displayed
for future reference.