BUN-6
Operator’s Manual – OPTI CCA-TS2
BUN/UREA
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
45
Range
5.5 to 103.2
Slope
1.01
Offset
-0.20
R
2
0.999
Sy|x
0.883
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Tietz, Burtis C.(Ed.),
Textbook of Clinical Chemistry and Molecular Diagnostics
, 4
th
Ed., (Elsevier
Saunders, 2006), p. 802.
3.
OPTI Medical.
Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0
20
40
60
80
100
120
0
20
40
60
80
100
120
B
U
N
-
TS2
BUN - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.20 + 1.01x)
Содержание CCA-TS2