Operator’s Manual – OPTI CCA-TS2
pH-B-3
ANALYTES
pH (Dry Sensor)
Chemical
Interferent
Concentration
pH Level
Interference
Methylene Blue
0.125 mM
7.170
NO
7.520
0.033
Sodium Chloride
20 mM
7.170
NO
7.520
NO
The following samples were identified as interfering with the dry pH sensor in the interference study
performed for the OPTI LION 510(k) submission.
Interferent
Test Level
Change
Sodium Bisulphate
11.5mM
-0.16
Phenylacetic Acid
10.0mM
-0.12
Methylene Blue
25mg/dL
Unstable
Fluorescein
25mg/dL
Unstable
Reproducibility
Controls
Within-run precision (S
wr
) and Total Precision (S
T
), were determined from 2 runs per day with 2
replicates per run over a period of 20 days following the CLSI guideline EP5-A2. Typical results for 3
different control levels are shown below:
Dry pH
OPTI Check Level 1
OPTI Check Level 2
OPTI Check Level 3
Days Run
20
20
20
Total Average
7.168
7.418
7.632
Within Run St. Dev. (S
wr
)
0.011
0.008
0.007
Within Run % CV
0.2%
0.1%
0.1%
Total Precision St. Dev. (S
T
)
0.015
0.011
0.011
Total % CV
0.2%
0.1%
0.1%
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on the OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
Dry pH
Whole Blood
Level 1
Level 2
Level 3
Average
7.174
7.352
7.558
St. Dev
0.011
0.019
0.012
%CV
0.16%
0.26%
0.17%
n
10
10
10