Lac-2
Operator’s Manual – OPTI CCA-TS2
LACTATE
ANALYTES
Standard Reference Cassette (SRC) Limit Values
LOW
NORMAL
HIGH
Units
1.00 ± 0.30
2.50 ± 0.50
5.00 ± 0.50
mmol/L
9.0 ± 2.7
22.5 ± 4.5
45.0 ± 4.5
mg/dL
Interferences
The following exogenous interferents were quantified in tonometered whole blood samples spiked with a
number of endogenous and exogenous chemicals and tested for interference following the CLSI guideline
EP7-A2, showing interferences to dyes which typically have short half-lives within the body before being
metabolized by the liver. This testing was done on the OPTI CCA-TS platform, with a standard OPTI
style Lactate sensor. No changes were made to the sensor or the measurement principle, so the
interference of the OPTI CCA-TS2 is equivalent to the OPTI CCA-TS.
Chemical
Test level
Concentration
d max
Lactate Level
mM
Interference
Acetaminophen
1.66 mM
0.2mM
2
NO
5
NO
Acetylsalicylic acid
3.33 mM
0.2mM
2
NO
5
NO
Ascorbic acid
0.23 mM
0.2mM
2
NO
5
NO
B-Hydroxybutyric acid
16.03 mM
0.2mM
2
NO
5
NO
Bilirubin
0.26 mM
0.2mM
2
NO
5
NO
Cardiogreen
0.0065 mM
0.2mM
2
0.4 mM
5
1.1 mM
Cystein
6.41 mM
0.2mM
2
NO
5
NO
Ethanol
86.8 mM
0.2mM
2
NO
5
NO
Evans blue
0.0104 mM
0.2mM
2
NO
5
NO
Glycolic acid
10 mM
0.2mM
2
1.4 mM
5
1.0 mM
Halothane
0.759 mM
0.2mM
2
NO
5
NO
Ibuprofen
2.43 mM
0.2mM
2
NO
5
NO