Operator’s Manual – OPTI CCA-TS2
pH-3
ANALYTES
pH
Whole Blood
Within-run precision in whole blood samples was evaluated at 3 different concentrations on one OPTI
CCA-TS2. The table below shows the Averages, Standard Deviations and % CV calculated from 10
samples for each level.
pH
Whole Blood
Level 1
Level 2
Level 3
Average
7.134
7.341
7.518
St. Dev
0.004
0.006
0.009
%CV
0.05%
0.08%
0.13%
n
10
10
10
Linearity
Wherever possible, linearity for the OPTI CCA-TS2 measurement has been established against reference
materials or methods. The linearity of the pH sensor on the OPTI CCA-TS2 has been established versus
the same sensor on the OPTI CCA-TS using whole blood samples tonometered with different %CO
2
gas
mixtures to establish the correlation.
Parameter and Specification
Value
Number of Samples
81
Slope
0.98
Offset
0.16
R
2
1.000
Sy|x
0.009