Ca-6
Operator’s Manual – OPTI CCA-TS2
CALCIUM
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
54
Range
0.28 to 2.12
Slope
1.00
Offset
0.00
R
2
0.999
Sy|x
0.013
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2.
Burritt MF, Pierides AM, Offord KP:
Comparative studies of total and ionized serum calcium values
in normal subjects and in patients with renal disorders.
Mayo Clinic proc. 55:606, 1980.
3. Tsien R,
New Calcium Indicators and Buffers with High Selectivity Against Magnesium and Protons
,
Biochemistry 19, p.2396-2404, 1980
.
4. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
2.2
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
2.2
C
a
-T
S2
Ca - TS
Scatter Plot with Fit
Identity
Linear fit
(0.00 + 1.00x)