5-10
Operator’s Manual – OPTI CCA-TS2
5 SAMPLE HANDLING AND PATIENT TESTING
•
When
using capillary tubes, push the tube
firmly into the fillport (Fig. 5-13).
WARNING:
Do not inject the sample!
It will be aspirated automatically.
•
The sample will then be aspirated
and
measured
(Fig. 5-14)
.
During the measurement, the status light is
blinking and a progress bar is displayed.
•
Do not open the cover of the sample
measurement chamber during the
measurement. If you do, the cassette and the
sample must be discarded.
15. You can again enter patient information
while measurement is in progress by pressing
<Patient Info>
(Fig. 5-14). Please follow the
steps for patient data entry described under
calibration on p. 5-8.
When the analysis is completed, the status light
stops blinking and the instrument alerts you that the
measurement has been completed with a “beep”.
At this time you may continue entering or editing the
patient information until you have completed it.
If the screen has not been touched for approximately
three (3) minutes, the
<Measurement Results>
will automatically be displayed (Fig. 5-15).
NOTE: The printout will start automatically when
the first results are displayed. This feature
may be turned off in setup (See Section
3.2.4.1).
NOTE: If ASTM, POCT1, or removable media is
enabled, the results will not be printed until
the user exits the measurement process.
The data will also be exported when it is
printed. Additional information on printing
reports and exporting data can be found in
Chapter 6, Data Management.
Fig. 5-15 Measurement results
Fig. 5-14 Sample Measurement
Fig. 5-13 Attach Capillary