Operator’s Manual – OPTI CCA-TS2
PCO2-5
ANALYTES
P
CO
2
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specification
Value
Number of Samples
117
Range
14.3 to 198.9
Slope
0.97
Offset
0.87
R
2
0.999
Sy|x
1.114
References
1. Tietz, Norbert W., Ed.,
Clinical Guide to Laboratory Tests
, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Vurek GG, Feustel PJ, Severinghaus JW,
A Fiber Optic
P
CO
2
Sensor
, Ann.Biomed.Eng. 11, p.499,
1983.
3. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0
20
40
60
80
100
120
140
160
180
200
0
20
40
60
80
100
120
140
160
180
200
PC
O
2
-T
S2
PCO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(0.87 + 0.97x)