8-32
Operator’s Manual – OPTI CCA-TS2
8 DIAGNOSTICS AND TROUBLESHOOTING
To set up the B-Lac cassettes, you have to run up
to 8 ampoules of OPTI Check level 2 (Fig. 8-38).
Please make sure you have at least 8 ampoules of the
same lot of OPTI Check and 8 B-Lac cassettes of the
same lot on hand before starting setup.
•
Press
to run the first sample.
Refer to section 4.5.2.1 for instructions on
running controls.
•
After each level of OPTI Check is run, you
will return to this screen (Fig. 8-39). You will
not receive a results screen or printout until the
setup procedure is complete. Continue to run the
OPTI check with the same lot of cassettes and
OPTI check until prompted. You may run from 5
to 8 OPTI checks.
•
After running the required number of samples,
the lactate parameter is enabled (Fig. 8-40).
•
Press
to exit the menu.
You will receive a B-Lac setup report once
complete.
After completing this procedure, OPTI Medical
recommends that you run two levels of OPTI Check
using B-Lac cassettes to verify performance.
Fig. 8-39 Run Controls
Fig. 8-40 B-Lac enabled
Fig. 8-38 Run Controls