Operator’s Manual – OPTI CCA-TS2
Na-5
ANALYTES
SODIUM
Correlation to Other Methods
3
OPTI CCA-TS vs other Electrolyte Instruments on Whole Blood and Serum in a typical
setting
Excess blood aliquots from specimens collected for blood gas analyses were analyzed by both traditional
and non-traditional operators of measurement equipment in hospital laboratories. The sample was
analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for
these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept
Correlation
Coefficient
Sy.x
Range
n
Analyzer A (whole blood)
0.6500
50.15
0.5721
2.21
126 to 149
103
Analyzer B (whole blood)
0.9313
9.34
0.9180
1.95
129 to 156
173
Analyzer C (whole Blood) 1.084 ± 0.226 -14.929 ± 3.176
0.9784
1.826
128 to 174
105
Analyzer D (whole blood) 1.080 ± 0.021
-6.382 ± 2.855
0.9678
2.007
117 to 163
174
Analyzer E (serum)
0.873
15.49
0.8911
1.77
128 to 149
68
Analyzer F (serum)
1.025
-4.57
0.9376
1.57
127 to 148
102
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.