Operator’s Manual – OPTI CCA-TS2
BUN-5
ANALYTES
BUN/UREA
Correlation to Other Methods
3
OPTI CCA-TS vs other BUN (urea) Instruments on Plasma and Serum in a typical setting
Excess sample aliquots from specimens collected for metabolyte analyses were analyzed by both
traditional and non-traditional operators of measurement equipment in hospital laboratories. The sample
was analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation
used for these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method*
Slope
Intercept Correlation
Coefficient
Sy.x
Range
n
Analyzer A (plasma) mg/dL
0.940
2.97
0.9975
1.056
8 to 89
68
Analyzer B (plasma) mg/dL
1.058
-3.04
0.9988
1.12
4 to 106
102
Analyzer C (serum) mg/dL
0.993
0.44
0.9953
1.00
6 to 65
47
Analyzer D (serum) mg/dL
0.971
-0.27
0.9822
0.98
5 to 42
50
* For more information on specific analyzers used, please contact OPTI Medical Technical Support.