2-6
Operator’s Manual – OPTI CCA-TS2
2 SETUP
6. Performing tHb Calibration
The tHb Calibrator Cassette should be run prior
to patient testing when first setting up your
analyzer. The tHb calibrator should then be run
every three months. Your OPTI CCA-TS2 will
remind you when the tHb calibration is due.
The tHb Calibrator Cassette can be found in
the storage compartment in the back of your
analyzer.
•
In the
<Ready>
display,
scan the bottom
bar code on the calibrator cassette package
by holding it 2-3 inches (5-8 cm) from the
bar code scanner located on the bottom
right-hand corner of the analyzer (Fig. 2-14).
NOTE: A tHb calibration can also be run
from the QC menu by pressing
<QC
Manager>QC>tHb Calibrator>
instead
of scanning the barcode in the
<Ready>
screen.
•
The red line from the bar code scanner
should cover the entire bar code.
•
A beep and a green status light indicates a
valid bar code.
•
If
<Non Secure User ID Entry>
is
enabled in the security settings (see Section
3.2.3), you will be asked to enter
the user ID
(Fig. 2-15). Depending on security settings,
user access to running Hb calibrators may be
restricted.
NOTE: Bar-coded user IDs may be entered from this
screen using the bar code scanner.
•
A warning will be displayed the first time a
new tHb Calibrator lot is used (Fig. 2-16).
Press
<Continue>
.
Fig. 2-15 Enter User ID
Fig. 2-14 Scan Bar Code
Fig. 2-16 New Calibrator